| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.172 | 2.077 |
| 2500.00 | 1.602 | 1.507 |
| 1250.00 | 1.113 | 1.018 |
| 625.00 | 0.863 | 0.768 |
| 312.50 | 0.529 | 0.434 |
| 156.25 | 0.336 | 0.241 |
| 78.13 | 0.228 | 0.133 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-99% | 93% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 93-105% | 82-101% | 91-103% |
| EDTA plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% |
| Heparin plasma(n=5) | 83-97% | 81-103% | 89-101% | 81-98% |