| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.951 | 1.849 |
| 10.00 | 1.536 | 1.434 |
| 5.00 | 1.279 | 1.177 |
| 2.50 | 0.896 | 0.794 |
| 1.25 | 0.507 | 0.405 |
| 0.63 | 0.388 | 0.286 |
| 0.32 | 0.226 | 0.124 |
| 0.00 | 0.102 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 80-92% | 86% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
| EDTA plasma(n=5) | 93-104% | 83-96% | 95-103% | 88-95% |
| Heparin plasma(n=5) | 86-94% | 93-102% | 87-96% | 92-105% |